TRANSMITTER SPECIFICATIONS
PROBE SPECIFICATIONS
PERFORMANCE
A complete measuring system consists of:
1. Jupiter 200: transmitter and probe (MLI or cages as shown in this bulletin are not included).
2. OPTION: secondary float for interface applications (specify S.G. for lower liquid).
3. OPTION: ATLAS
®
. Magnetic level indicator for use with Jupiter 200, external mount model. Consult bulletin 46-138.
4. Free of charge: Magnetrol master C.D. with Jupiter 200 DTM (PACT
war e
®
). Order code: 090-BE59-200 (included in each
order).
MODEL IDENTIFICATION
Description Specification
Accuracy ± 0,4 mm (0.015")
Repeatability ± 0,005 % of full span or 0,13 mm (0.005") – whichever is greater
Linearity ± 0,020 % of full span or 0,79 mm (0.031") – whichever is greater
Max fill / drain rate 9 m/min (30 ft/min)
Response Time < 0,1 second
Warm-up Time < 5 seconds
Ambient Temp. -20 °C to +70 °C (-5 °F to +160 °F)
Humidity 0-99 %, non-condensing
Electromagnetic Compatibility Meets CE requirements (EN-61000-6-4, EN 61000-6-2)
Description Specification
Materials Probe 316/316L (1.4401/1.4404) standard or electropolished finish
Hastelloy C
®
(2.4819) or Monel
®
(2.4360)
Float 316 (1.4401), 316L (1.4404), Titanium or Hastelloy C
®
(2.4819)
Process seal None, welded construction
Probe diameter 16 mm (0.63")
Float diameter
(for direct insertion models only)
Min 47 mm (1.85") – max 65 mm (2.55")
see partnumber on page 7
Probe length Min 15 cm (6") – max 570 cm (224") selectable per 1 cm increments
Dead band Upper Direct insertion model: 50 mm (2").
External mount model: depending configuration.
Bottom 50 mm (2") – for units with standard electronics and single float
152 mm (6") – for units with standard electronics and dual floats
Inactive zone - bottom 127 mm (5") – for units with SIL enhanced electronics
Process
temperature
Direct insertion -40 °C to +95 °C (-40 °F to +200 °F) – standard probe
-40
°
C to +260
°
C (-40 °F to +500 °F) – high temperature probe
External mount -40
°C to +120 °C (-40 °F to +250 °F) – standard
-196 °C to +455 °C (-320 °F to +850 °F) – with factory insulated MLI
Process
Pressure
316 and 316L Max 22,8 bar @ +40 °C (330 psig @ 100 °F)
Titanium Max 26,2 bar @ +40 °C (380 psig @ 100 °F)
Hastelloy C Max 18,6 bar @ +40 °C (270 psig @ 100 °F)
Vacuum service Full vacuum
Consult factory for higher pressure (custom float).